Electronic Supplementary Material to the article: "Effects Of Electromagnetic Interference From Radio Frequency IDentification On Critical Care Equipment"

Logo JAMA© MICU-department of the Academic Medical Center, Amsterdam, the Netherlands, 2008. E.J.vanLieshout@amc.uva.nl

All rights reserved. Permission in writing is required before reproducing any content of the publications of the MICU-department of the Academic Medical Center. Those requests that are approved will be issued a grant of permission by mail.

type of device interference signal  classification details
(distance in cm) (type)
infusion / syringe pumps
Graseby 3500 100 868MHz H stops working, acoustic alarm, error messages, resetting not possible.
Alaris IVAC 591 50 868MHz H stops working, acoustic alarm, error messages, after reset pump works normal again.
B.Braun Compact(S) 40 868MHz L beap heard from loudspeaker.
B.Braun Infusomat (P) 30 868MHz L rattling noise heard from loudspeaker.
B.Braun Infusomat (P) 5 125kHz H stops operating, acoustic alarm, error message.
Alaris IVAC 598 10 868MHz H stops operating, acoustic alarm, error messages.
B.Braun perfusor ft 10 868MHz H stops operating, acoustic alarm.
B.Braun perfusor fm 0,1 868MHz L lamp 'battery' switches off.
B.Braun Infusomat Space

no electromagnetic interference

external pacemakers    
Medtronic 5348 Single chamber 30 868MHz H on demand: pace pulses are given erroneously.
Medtronic 5348 Single chamber 30 125kHz H on demand: pace pulses are given erroneously.
Biotronik EDP20/A 25 868MHz H pace pulses are given erroneously.
Medtronic 5388 Dual chamber 10 868MHz H on demand: pace pulses are given erroneously in both atrium and ventricle.
Medtronic 5388 Dual chamber 5 125kHz H on demand: pace pulses are given erroneously in both atrium and ventricle.
mechanical ventilators
Hamilton Galileo 400 868MHz H ventilation triggers erroneously, higher ventilation frequency
Hamilton Galileo 20 125kHz S interference at flow curve at display.
Drager Evita 4 5 868MHz H stops operating, accoustic alarm, loss of settings, automatic restart.
Hamilton Raphael color & Siemens Servo 300A & screen 390

no electromagnetic interference

hemofiltration/dialysis devices    
Gambro AK200 ULTRA S  20 868MHz H stops operating (recoverable from operation panel), acoustic alarm.  can be continued. At 0 cm: stops working, sometimes acoustic alarm, treatment stops, reset
B.Braun Diapact CRRT 10 868MHz H stops operating (recoverable from operation panel), acoustic alarm.
pacemaker programmers    
Guidant Zoom Latitude programming system 3120 600 868MHz S interference in atrial & ventricular electrocardiogram (EGM)
Guidant Zoom Latitude programming system 3120 25 125kHz H active tag, solely : interference in atrial & ventricular electrocardiogram (EGM)
Medtronic 2090 The Programmer 100 868MHz H communication with ICD disturbed, LED-bar on programming head disrupted. 
Medtronic 2090 The Programmer 200 125kHz H communication with ICD disturbed, LED-bar on programming head disrupted.
intra-aortic balloon pumps    
Datascope System 97 50 868MHz H stops operating, acoustic alarm, settings lost.
Arrow Acat 1 plus & Arrow AutoCat2Wave.

no electromagnetic interference

fluid warmer    
Surgical company Fluido 50 868MHz H resets automatically and resumes operating normally when signal is removed.
cardiopulmonary bypass device    
Sarns 9000 10 868MHz H cardioplegy pump into band skid mode, no change in rotational speed.
autologous blood recovery device    
Haemonetics HaemoLite2Plus 5 868MHz H change in rotational speed
anaesthesia devices    
Agilent/HP CM-S Monitor with ECG module M1001A/B 600 868MHz L interference in ECG.
Agilent/HP CM-S Monitor with ECG module M1001A/B 50 125kHz L interference in ECG.
GE S/5 Avance, GE Aestiva/5, Philips Airway Gases M1026B

no electromagnetic interference

defibrillators    
Philips Heartstart 600 868MHz L interference noise from loudspeaker.
Medtronic lifepak 20 5 868MHz L interference noise from loudspeaker.
Medtronic lifepak 12

no electromagnetic interference

12 leads electrocardiogram device    
Marquette MAC5000 12 leads ECG  250 868MHz L interference in ECG.
Marquette MAC5000 12 leads ECG  25 125kHz L interference in ECG.
monitors    
Philips Intellivue MP30 with module M3001A 50 868MHz L noice from loudspeaker.
PhilipsIntellivue MP50 & Philips Intellivue MP90

no electromagnetic interference

IC bed    
Hill-Rom Avant Guard & Hill-Rom Total Care

no electromagnetic interference

operating table    
Maquet Alphamaquet

no electromagnetic interference

hyper/hypothermia device    
Sarns/Terumo TCM II

no electromagnetic interference

vacuum pump    
Laerdal LSU

no electromagnetic interference